The GDS-DOZONE Ozone Standard Gas Generator meets the technical requirements for ozone standard gas generation devices specified in the "JJG1077-2012 Verification Regulation of Ozone Gas Analyzers". It can generate high-concentration and low-concentration ozone standard gases. The high-concentration ozone standard gas generator uses corona discharge principle and can produce ozone standard gas with concentrations up to 400ppm; the low-concentration ozone standard gas generator uses ultraviolet oxygen photolysis principle to produce ozone concentrations up to 5ppm. It is suitable for calibration of ozone analyzers and transfer standards.
Detailed Introduction
GDS-DOZONE Ozone Standard Gas Generator The GDS-DOZONE Ozone Standard Gas Generator meets the technical requirements for ozone standard gas generation devices specified in the "JJG1077-2012 Verification Regulation of Ozone Gas Analyzers". It can generate high-concentration and low-concentration ozone standard gases. The high-concentration ozone standard gas generator uses corona discharge principle and can produce ozone standard gas with concentrations up to 400ppm; the low-concentration ozone standard gas generator uses ultraviolet oxygen photolysis principle to produce ozone concentrations up to 5ppm. It is suitable for calibration of ozone analyzers and transfer standards.
The GDS-DOZONE Ozone Standard Gas Generator integrates a dual-beam UV spectrophotometric ozone analysis detection system, which analyzes and detects the concentration of ozone gas output by the generator in real time. The instrument features a 7-inch LCD touch screen, allowing users to set the output of ozone standard gas at specified concentrations without being affected by ambient temperature, pressure, or humidity. The output flow rate can reach 6L/Min. Users can set the desired ozone concentration using the output concentration setting function key on the operation interface, and click the start button to generate standard gas of the specified concentration in real time. Users can directly connect the output of the ozone calibrator to the inlet of an ozone analyzer or an ozone alarm.
Product Features:
Response time less than 25 seconds; Ozone generator flow rate up to 6L/min; Equipped with temperature and pressure correction; Touch screen operation; Built-in high-concentration ozone generator using corona discharge method; Built-in low-concentration ozone generator using ultraviolet oxygen photolysis method;
Compact and durable instrument case;
Easy-to-use menu; Technical Parameters High-concentration ozone generation principle Corona discharge High-concentration ozone output range 0-400ppm High-concentration ozone output maximum error <5% Low-concentration ozone generation principle Ultraviolet oxygen photolysis Low-concentration ozone output range 0-5ppm Low-concentration ozone output maximum error <2% Output concentration stabilization time High concentration < 300s
Low concentration < 120s Output flow rate 6L/min Zero noise 0.8ppb (60-second average time) Detection limit 2ppb Gas source requirements Zero-grade air >10L/min(@0.3MP) Oxygen >2L/min(@0.15MP) Operating temperature 0-45 ℃ Power supply AC220V (built-in lithium battery) Dimensions 330(L)*179(W)*260(H) Product Applications: Calibrate ozone detectors, ozone alarms, etc. Study the effects of ozone on plants; this application requires a calibrated ozone calibration source. This transfer standard instrument is particularly useful for maintaining the calibration of ozone instrument networks.